DESCRIPTION

The Model 610 and 650 Current Shunts are specifically designed to operate with the Model 6000A Precision Phase Meter.   These shunts are frequency tuned to have negligible phase shift (less than ± 5mº) at 50/60Hz making them ideal for line frequency applications.  They can also be idealized for other frequencies as specified by the user.  Consult factory for details.

SPECIFICATIONS

Model Current Range

Frequency

Phase Shift
610 1A to 10A

5Hz to 1kHz 
1kHz to 100kHz

Less than 5mº
5mº+50mº(fkHz -1)

650 10A to 50A

5Hz to 100Hz
100Hz to 500Hz
500Hz to 1kHz

Less than 5mº
Less than 10mº
50mº

The Model 610 and 650 are designed to drive the high input impedance of the Model 6000A  phase meter.  Any other meter used with these shunts must have an input resistance  greater than 1Mohm in parallel with not more than 100pF to avoid additional phase error.